FTIR Gasmet In-Situ 5.0

As the first company in the world, Gasmet Technologies introduces an In Situ FTIR gas analyser for stack emissions monitoring. The In Situ analyser consists of a sample cell inserted into the stack or duct, heated steel mounting flange and a rugged interferometer unit in an IP65 enclosure that is directly attached to the flange.

 

The Gasmet In Situ analyser offers a cost-effective alternative for extractive systems in continuous emissions monitoring applications where multiple components are monitored. By combining powerful FTIR technology with rugged in-stack measurement cell, the Gasmet In Situ offers increased analytical capability and reduced cross interference effects when compared with other in-situ analysers.

The Gasmet™ In-Situ analyser consists of a sample cell inserted into the stack or duct, heated steel mounting flange and the rugged GICCOR™ interferometer unit in an IP65 enclosure that is directly attached to the flange. Gasmet™ In-Situ can generate direct 4-20 mA outputs for the measurement results, or alternatively it can be connected to an external PC for further data analysis and logging features.

The Gasmet™ In-Situ offers a cost-effective alternative for extractive systems in continuous emissions & process monitoring where multiple compounds are monitored. Typically, concentrations of H2O, CO2, CO, NO, NO2, N2O, SO2, HCl, HF, NH3, CH4 are continuously monitored with a single Gasmet™ In-Situ. In addition, different VOC’s can be measured as individual compounds. There is no need to do any span calibrations; just daily zero calibration with dry instrument air.

By combining powerful FTIR technology with rugged in-stack measurement cell, the Gasmet™ In-Situ offers increased analytical capability and reduced cross-interference effects when compared with other in-situ analysers.

 

In-Situ Multicomponent FTIR Gas Analyser

 

GASMET FTIR In-Situ SERIES includes industrial multicomponent gas analysers for continuous monitoring applications. The GASMET In-Situ 5.0 incorporates a Fourier Transform Infrared spectrometer, a sample cell which is directly inserted into the sample gas flow, and signal processing electronics. The analyser is fully equipped for fixed installations and it offers versatility and high performance for a wide range of applications. Standard In-Situ 5.0 analyzer includes analog outputs 4-20 mA for measuring data and relay contacts for alarms. As an option the In-Situ analyser can be equipped with an external computer unit. In this case the In-Situ employs the same Calcmet user interface as an extractive GASMET CEMS. The sample must be non-condensing and the upper limit for sample temperature is 250 oC. Sample cell absorption path length is 5.0 meters. Insertion depth of the sample probe is

590 mm and the flange diameter is 240 mm.

 

General parameters

Measuring principle: Fourier Transform Infrared, FTIR

Performance: simultaneous analysis of up to 50 gas compounds

Response time, T90: typically < 120 s, depending on the gas flow and measurement time

Operating temperature: -30 – 40 °C non condensing

Storage temperature: -30 – 60 °C, non condensing

Power supply: 100-115 or 230 V / 50 -60 Hz

Power consumption: 500 W max.

Instrument air: Dew point –20 oC, oil free

Max. air consumption:

120 l/min for vortex cooling (cont.)

100 l/min for zero calibration/purge

(15 minutes at 24 hour intervals)

 

Spectrometer

Resolution: 8 cm-1

Scan frequency: 10 scans / s

Detector: Peltier cooled MCT

Source: SiC, 1550 K

Beamsplitter: ZnSe

Window material: ZnSe

Wavenumber range: 900 - 4 200 cm-1

 

Sample Cell

Structure: Multi-pass, fixed path length 6.0 m

Probe body material: AISI 316 steel

Mirror material: Rhodium coated aluminium

Mirror coating: Reflective gold coating

Gaskets: Viton® O-rings

Window material: BaF2

Insertion depth 589 mm

Mounting flange 240 mm diameter

Measuring parameters

Zero point calibration: 24 hours, calibration with instrument air

Zero point drift: < 2 % of measuring range per zero point calibration interval

Sensitivity drift: none

Linearity deviation: < 2 % of measuring range

Temperature drifts: < 2 % of measuring range per 10 K temperature change

Pressure influence: 1 % change of measuring value for 1 % sample pressure change. Sample pressure changes are measured and compensated

 

Sample Gas Conditions

Gas temperature: Up to 250 oC, non condensing

 

Electrical Connectors

A/D Converter: Dynamic range 95 dB

Signal Processor: 2 32-bit floating point DSP’s 120 MFLOPS

Digital Interface: Bayonet locking 10-pole circular connector (MIL-C-5015) for serial and relay output.

Measuring data: 4-20 mA, max 8-channel

Alarm Relays: Function, alarm, service relays

Power connection: Terminal connectors for L + N + PE wires

 

Optional Computer Unit

Analyzer is connected to an external computer via serial cable. The external computer controls the GASMET In-Situ analyzer

Minimum Configuration: 512 MB Memory, > 40 GB hard disk, > 2.4 GHz Intel Pentium IV Processor, Modem, Network card

Operating system: Windows Xp operating system

Software included: Calcmet for Windows

Watchdog support: Included

Relay contacts (status): Function Fault

Service required

Sampling Alarm

Relay contacts (valves): Sample gas

Zero gas

Span gas 1

Mouse: Included

Keyboard: Not Included

Display: TFT display

Digital Interface RS-232 cable to analyzer, max

10m

Remote control: Built-in modem/network card and

PC Anywhere support

Options

Adapters for flanges larger than 240 mm

RS-422/485 interface instead of RS-232 for longer distance,

max 1 km.

Enclosure

Material: Aluminium / Steel /

Polyethylethylketone (PEEK)

Dimensions (mm): 1018 x 390 x 250

Weight: 30 kg

CE - Label: According to EMI guideline